Electromagnetic flowmeter detection method and technical solution for online verification
Sep 04, 2018

 instruments and tools required for the detection of the electromagnetic flowmeter itself are: a multimeter, a process calibrator, a 500V megohm meter and common tools.

1.  instruments and tools required for the detection of the electromagnetic flowmeter itself are: a multimeter, a SUPCON_X319 process calibrator, a 500V megohm meter and common tools.

2. Detection of electrode assembly and measuring tube assembly
Measuring the contact resistance between the electrode and the liquid (in the presence of a medium, the line numbers "SGND" and "SIG1" and "SGND" and "SIG2" to determine whether there is a short-circuit between the electrode and the ground, the short-circuit between the poles, indirectly assess the electrode, lining surface general situation