Plug-in electromagnetic flowmeter technical parameters
Sep 05, 2018

Plug-in electromagnetic flowmeter technical parameters

 

Path: DN200-3000

 

Electrode material: Mo2Ti, HC, HB, Pt, Ta, Ti

 

Measuring medium: conductive liquid

 

Dielectric conductivity: >5μs/cm (water >20μs/cm)

 

Accuracy level: ±1.0%RS

 

Repeatability: ±0.1%

 

Flow rate range: 0-10m/s

 

Maximum medium temperature: integrated type ≤90°C split type ≤180°C

 

Ambient temperature: -25°C-65°C

 

Working pressure: 0.6Mpa-1.6Mpa

 

Protection level: IP65, IP67, IP68 (optional)

 

Output signal: 4-20mA current output, 0-1kHz frequency output, unit volume pulse output

 

Power: 220VAC, 24VAC, 3.6V battery powered

 

Faults and troubleshooting of plug-in electromagnetic flowmeter

 

Fault 1: There is no output current after the power is turned on. Cause: The power is not turned on.

 

Removal method 1. Check if the fuse is blown 2. Check the power circuit wiring.

 

Fault 2: The current output is zero (4 mA) after the power is turned on. Cause: 1. The direction of the sensor mark is opposite to the flow of the fluid. 2. The converter excitation power line is connected incorrectly. 3. The input signal lines E1 and E2 are reversed.

 

Exclusion method: 1. Adjust the direction of the sensor to match the direction of the fluid. 2. Rewire the cable in the correct way. 3. Replace the signal lines E1 and E2.

 

Fault 3: The output signal fluctuates too much. Cause: 1. There is gas at the sensor electrode, which causes the electrode to be in poor contact with the measured medium. 2. There is accumulation on the surface of the electrode, and the electrode is in poor contact with the medium. 3. The media flow actually fluctuates greatly. 4. Poor grounding.

 

Discharge method: 1. Correct installation to eliminate the influence of gas in the pipeline. 2. Clean the electrode surface. 3. Increase the meter damping time.


  • Newsletter
  • Categories
  • Contact Us
    Beijing Gallop Technology Co.,Ltd
    Add:Room410 Tower C, Jiahua Buiding NO.9 Shangdi 3 Street Haidian District, Beijing
    Tel:+86-010-51269778
    Fax:+86-010-62967939
    Mobile:+86-13858586115
    E-mail:13858586115@163.com
  • QR Code
  • Copyright © Beijing Gallop Technology Co.,Ltd All Rights Reserved.